Service checks

The annual service check at the KaVo factory includes a test of all functions, components, software updates, calibration and "MPG" testing according to the Medical Devices Act.

 

Scope of services KaVo Service Check INTRAsurg

  • Surgical motor: Cleaning and functional test
  • Surgical hoses: Cleaning and functional test
  • Motor electronics: Functional test and adjustment
  • Software update: Update to most current software
  • Hose pump: Check and power adjustment of hose pump 1 and, if applicable, hose pump 2 
  • Optional air connector for INTRAsurg 1000: Check and adjustment of air pressure parameters
  • Data transmission for the INTRAsurg 1000: Check of data output to PC
  • Foot-operation element: Check of mechanical and electrical functions
  • Calibration: Calibration of speed and torque
  • Implementation of the obligatory annual safety checks according to VDE 0751/1

 

Scope of services KaVo Service Check STERIclave

  • Cleaning: A cleaning cycle will be carried out. Pressure chamber and fresh water tank will be cleaned. Dirt and dust will be removed from cooling radiator
  • Vacuum pump: Check of performance and function
  • Parts subject to wear: Replacement, in due time, of various parts that are subject to wear, e.g.: door gasket, filter, non-return valves
  • Calibration: Check if the temperature and pressure profile is accurate
  • Validation in accordance with the DIN EN 13060 standard
  • Safety checks in accordance with VDE 0701/1

 

Scope of services KaVo Service Check GENTLEray

  • Laser handpiece: Check of laser coupling and function
  • Laser fibre: Check of energy transmission
  • Software update: Update to most current software
  • Calibration: Check and adjustment of accurate laser output power
  • Laser protection facilities: Check of laser safety facilities
  • Safety checks: Checks in accordance with VDE 0751/1

 

Scope of services KaVo Service Check HealOzone

  • HealOzone handpiece and hose: Functional test of handpiece and hose, replacement of O-rings
  • Software update: Update to most current software
  • Ozone destruct system: Check, and replacement if needed, of activated carbon cartridge
  • Moisture trap: Replacement of the moisture trap
  • Air dryer: Check of the air dryer for damage and function. Replacement of the drying beads
  • Calibration: Check and adjustment of the accurate ozone concentration  
  • Safety checks: Checks in accordance with VDE 0751/1

 

Scope of service KaVo MASTERsurg/EXPERTsurg

  • Surgical motor: Cleaning and functional test
  • Check the current consumption
  • Surgical hoses: Cleaning and functional test
  • Motor electronics: Functional test and adjustment
  • Software update: Update to most current software
  • Hose pump: Check and power adjustment of hose pump
  • Check sensor coolant hose
  • Check display and touch function
  • Foot-operation element: Check of mechanical and electrical functions
  • Calibration: Calibration of speed and torque
  • Implementation of the safety checks according to VDE 0751/1

 

Scope of service KaVo EXPERTlase/MASTERlase

  • Laser handpiece: Check of laser coupling and function
  • Laser fibre: Check of energy transmission
  • Check of pilot beam
  • Software update: Update to most current software
  • Calibration: Check and adjustment of accurate laser output power
  • Laser diode: Performance check
  • Laser protection facilities: Check of laser safety facilities
  • Check of the obligatory annual safety checks according to VDE 0751/1

 

Scope of service KaVo DIAGNOcam

  • Clean the optics
  • Check the latest technical stand and rebuild if necessary
  • Device detection/ Check control buttons
  • Check tip functions – Switching capacity, light guide, cracks, splice
  • Check heating of the optics
  • Check image quality with boundary parts
  • Laser power check and adjust if necessary
  • Timer check and adjust if necessary
  • Timeout check and adjust if necessary
  • Check tray switch